Open access
Research Article
23 May 2017

Comparison of ePlex Respiratory Pathogen Panel with Laboratory-Developed Real-Time PCR Assays for Detection of Respiratory Pathogens

ABSTRACT

Infections of the respiratory tract can be caused by a diversity of pathogens, both viral and bacterial. Rapid microbiological diagnosis ensures appropriate antimicrobial therapy as well as effective implementation of isolation precautions. The ePlex respiratory pathogen panel (RP panel) is a novel molecular biology-based assay, developed by GenMark Diagnostics, Inc. (Carlsbad, CA), to be performed within a single cartridge for the diagnosis of 25 respiratory pathogens (viral and bacterial). The objective of this study was to compare the performance of the RP panel with those of laboratory-developed real-time PCR assays, using a variety of previously collected clinical respiratory specimens. A total of 343 clinical specimens as well as 29 external quality assessment (EQA) specimens and 2 different Middle East respiratory syndrome coronavirus isolates have been assessed in this study. The RP panel showed an agreement of 97.4% with the real-time PCR assay regarding 464 pathogens found in the clinical specimens. All pathogens present in clinical samples and EQA samples with a threshold cycle (CT) value of <30 were detected correctly using the RP panel. The RP panel detected 17 additional pathogens, 7 of which could be confirmed by discrepant testing. In conclusion, this study shows excellent performance of the RP panel in comparison to real-time PCR assays for the detection of respiratory pathogens. The ePlex system provided a large amount of useful diagnostic data within a short time frame, with minimal hands-on time, and can therefore potentially be used for rapid diagnostic sample-to-answer testing, in either a laboratory or a decentralized setting.

INTRODUCTION

Infections of the upper and lower respiratory tract can be caused by a diversity of pathogens, both viral and bacterial. Community-acquired respiratory tract infections are a leading cause of hospitalization and responsible for substantial morbidity and mortality, especially in infants, the elderly, and immunocompromised patients. The etiological agent in such infections differs greatly according to season and age of patient, with highest prevalences being those of respiratory syncytial virus (RSV) in children and influenza virus in adults. Rapid microbiological diagnosis of a respiratory infection is important to ensure appropriate antimicrobial therapy and for the effective implementation of isolation precautions (1).
In the last decade, many conventional diagnostic methods such as culture and antigen detection assays have been replaced by molecular assays for diagnosing respiratory tract infections. Multiplex real-time PCR assays have been developed and implemented for routine diagnostic application, detecting a wide variety of pathogens (27). These assays have shown high sensitivity and specificity, but the limited number of fluorophores that can be used per reaction resulted in the need to run several real-time PCR assays to cover a broad range of relevant pathogens. Commercial assays using multiplex ligation-dependent probe amplification (MLPA), a dual priming oligonucleotide system (DPO), or a microarray technology were developed to overcome this problem and are able to detect up to 19 viruses simultaneously (8, 9). All applications mentioned require nucleic acid extraction prior to amplification. For routine diagnostics, these methods are most suited for batch-wise testing, with a turnaround time of ∼6 to 8 h. To decrease the time to result and enable random access testing, syndromic diagnostic assays have been developed. These assays combine nucleic acid extraction, amplification, and detection in a single cartridge per sample and are suitable for decentralized or even point-of-care testing (POCT) with a time to result of <2 h.
A novel rapid diagnostic, cartridge-based assay for the detection of respiratory tract pathogens using the ePlex system (Fig. 1) was developed by GenMark Diagnostics, Inc. (Carlsbad, CA). The ePlex respiratory pathogen panel (RP panel) is based on electrowetting technology, a digital microfluidic technology by which droplets of sample and reagents can be moved efficiently within a network of contiguous electrodes in the ePlex cartridge, enabling rapid thermal cycling for a short time to result. Following nucleic acid extraction and amplification, detection and identification are performed using the eSensor detection technology (Fig. 2), as previously applied in the XT-8 system (10).
FIG 1
FIG 1 ePlex system (A) and the corresponding cartridge of the ePlex respiratory pathogen panel (B).
FIG 2
FIG 2 Principle of the eSensor detection technology. Amplified sequences of the targeted pathogens are detected electrochemically using a complementary pathogen-specific signal probe tagged with ferrocene, a reducing agent. The hybridized molecule is then exposed to another sequence-specific probe that is bound to a solid phase, which is a gold electrode (A). Upon binding of the two molecules, the ferrocene comes in close proximity to the gold electrode, where an electron transfer that can be measured using GenMark's eSensor technology on the ePlex system can occur (B).
In the current study, the performance of the syndromic RP panel was compared to those of laboratory-developed real-time PCR assays, using clinical specimens previously submitted for diagnosis of respiratory pathogens.

RESULTS

The 323 positive clinical specimens contained a total of 464 respiratory pathogens as detected by laboratory-developed real-time PCR assays (Table 1). As shown in Table 2, the 57 nonnasopharyngeal (non-NPS) specimens comprised 69 of the total 464 respiratory pathogens. Testing all samples with the RP panel resulted in an overall agreement for 452 (97.4%) targets from 311 specimens, prior to discrepant analysis. Of the specimens containing a single pathogen, the detected targets were concordant in 209/217 specimens. For samples with coinfection, the same pathogens could be identified in 77/81, 22/22, and 3/3 in the case of 2, 3, and 4 pathogens present, respectively. Eight of 12 discordant targets (PCR+/RP) had a positive result with threshold cycle (CT) values of >35 (Fig. 3). Retesting with a third assay confirmed 10 of 12 real-time PCR-positive targets being human bocavirus (hBoV; n = 3), rhinovirus (RV; n = 2), parainfluenza virus type 2 (PIV2; n = 1)), human coronavirus (hCoV) OC43 (n = 1), hCoV 229E (n = 1), hCoV HKU1 (n = 1), and human metapneumovirus (hMPV; n = 1). The two unresolved PCR+/RP results consisted of two hMPV-positive samples (CT values of 33.2 and 38.3).
TABLE 1
TABLE 1 Pathogens included in this study
RP panel targetNo. of results
Found in clinical specimensPCR+/RPPCR/RP+
Viral
    Adenovirus394
    Coronavirus
        229E711
        HKU1121
        NL6371
        OC43911
        MERS coronavirus
    Human bocavirus2731
    Human metapneumovirus283
    Human rhinovirus/enterovirusa134b2c6a
    Influenza virus
        Influenza A1d
            H15
            2009 H1N1181e
            H317
        Influenza B20
    Parainfluenza virus
        Type 111
        Type 2121
        Type 315
        Type 422
    Respiratory syncytial virus13d
        Type A43
        Type B24
Bacterial
    Bordetella pertussis6
    Chlamydophila pneumoniae0
    Legionella pneumophila6
    Mycoplasma pneumoniae8
a
No differentiation possible between rhinovirus and enterovirus.
b
One hundred seven rhinoviruses and 27 enteroviruses found by routine testing.
c
Both rhinovirus.
d
No further subtyping performed.
e
In the laboratory-developed test detected as influenza A virus.
TABLE 2
TABLE 2 Non-NPSa specimens included in this study
Specimen typeNo. includedPathogen (n)
Sputum21Human rhinovirus (10)
Legionella pneumophila (6)
Human metapneumovirus (3)
Respiratory syncytial virus (1)
Influenza A virus H1N1 (1)
Influenza A virus H3 (1)
Mycoplasma pneumoniae (1)
Bronchoalveolar lavage fluids16Human rhinovirus (12)
Human metapneumovirus (3)
Enterovirus (2)
Human bocavirus (2)
Respiratory syncytial virus (1)
Mycoplasma pneumoniae (1)
Adenovirus (1)
Throat swab10Human rhinovirus (7)
Respiratory syncytial virus (4)
Enterovirus (1)
Nasopharyngeal aspirate10Respiratory syncytial virus (7)
Human rhinovirus (3)
Enterovirus (1)
Adenovirus (1)
a
NPS, nasopharyngeal swab.
FIG 3
FIG 3 Pathogen concordance by CT value.
The RP panel yielded a positive result in 17 specimens, where the laboratory-developed test (LDT) remained negative (PCR/RP+), including 15 additional pathogens previously undetected by LDT in the 323 positive specimens and one influenza A H1N1 2009 virus that was detected as influenza A virus by LDT (Table 1). Seven of these 15 additional targets could be confirmed, including three of RV/enterovirus (EV) (all confirmed as RV), two of PIV4, and one each of hBoV and hCoV NL63.
One of the selected negative samples tested positive for human adenovirus (hAdV) in the RP panel but could not be confirmed by discrepant testing. All other negative specimens tested negative in the RP panel as well.
Both Middle East respiratory syndrome coronavirus (MERS-CoV) isolates could be detected by the RP panel. By testing a 10-fold dilution series of both isolates, it was shown that MERS-CoV with a CT value of <30 in the laboratory-developed real-time PCR assay could be detected using the RP panel, while detection with a CT value of >30 was achievable but was not reproducible in every instance.
Of the 12 specimens from the Quality Control for Molecular Diagnostics (QCMD) 2016 Respiratory II Pilot external quality assessment (EQA) study panel, 10 were detected in full agreement with the content as reported by QCMD (Table 3). The 2 false-negative tested specimens both contained hCoV NL63, of which one was a coinfection in an hMPV-positive sample. Both specimens had been tested with the laboratory-developed real-time PCR assay as well and were found positive for hCoV NL63, both with CT values of 37.4.
TABLE 3
TABLE 3 EQA samples included in the study
EQA sampleContentaRP panel
RESPII16-01Rhinovirus type 72 d1Rhinovirus/enterovirus
RESPII16-02Rhinovirus type 72 d2Rhinovirus/enterovirus
RESPII16-03NegativeNegative
RESPII16-04Adenovirus type 1Adenovirus
RESPII16-05Coronavirus OC43Coronavirus OC43
RESPII16-06 bHuman metapneumovirus and coronavirus NL63Human metapneumovirus
RESPII16-07Rhinovirus type 8Rhinovirus/enterovirus
RESPII16-08Parainfluenza virus type 3 and rhinovirus type 72Parainfluenza virus type 3 and rhinovirus/enterovirus
RESPII16-09Human metapneumovirusHuman metapneumovirus
RESPII16-10Coronavirus NL63 d1Coronavirus NL63
RESPII16-11Coronavirus NL63 d2Negative
RESPII16-12Parainfluenza virus type 1Parainfluenza virus type 1
Qnostics RSP-S01Influenza A virusInfluenza A virus
Qnostics RSP-S02Influenza B virusInfluenza B virus
Qnostics RSP-S03Respiratory syncytial virus type ARespiratory syncytial virus type A
Qnostics RSP-S04Respiratory syncytial virus type BRespiratory syncytial virus type B
Qnostics RSP-S05Adenovirus type 1Negative
Qnostics RSP-S06Coronavirus NL63Coronavirus NL63
Qnostics RSP-S07Coronavirus OC43Coronavirus OC43
Qnostics RSP-S08Parainfluenza virus type 1Parainfluenza virus type 1
Qnostics RSP-S09Parainfluenza virus type 3 and rhinovirusParainfluenza virus type 3 and rhinovirus/enterovirus
Qnostics RSP-S10Human metapneumovirusHuman metapneumovirus
Qnostics RSP-S11RhinovirusRhinovirus/enterovirus
Qnostics RSP-S12Bordetella pertussisBordetella pertussis
Qnostics RSP-S13 Chlamydophila pneumoniae Negative
Qnostics RSP-S14Mycoplasma pneumoniaeMycoplasma pneumoniae
Qnostics RSP-S15Legionella pneumophilaLegionella pneumophila
Qnostics RSP-S16Legionella pneumophilaLegionella pneumophila
Qnostics RSP-S17NegativeNegative
a
d1 and d2, different dilutions of the same virus, no information on the concentration provided.
b
Data in bold indicate samples that incorrectly detected the (full) content of the EQA sample.
The Qnostics evaluation panel consisted of 17 samples, including 15 different respiratory pathogens and one negative sample (Table 3). The RP panel detected 15 of the specimens in agreement with the content, whereas hAdV type 1 and Chlamydophila pneumoniae were not detected. Real-time PCR detection of these specimens was performed to confirm the presence of the respective pathogen in the specimen and was found positive for both hAdV (CT value of 31.4) and C. pneumoniae (CT value of 35.4).

DISCUSSION

The performance of the ePlex RP panel was assessed by retrospective testing of 343 clinical respiratory specimens (obtained in 2009 to 2016) comprising five different types of specimens. Although the RP panel had been CE in vitro diagnostic (CE-IVD) cleared for detection of respiratory pathogens from NPS swabs only, we included a range of alternate sample types that can be obtained and tested for respiratory pathogens in the diagnostic setting. By including a total of 57 respiratory non-NPS specimens with different pathogens (Table 2), it was shown that the RP panel was able to accurately detect the pathogen(s) in the different types of specimens, as the assay showed 100% concordance with LDT. For sputum samples, preprocessing with Sputasol was introduced after the initial 6 tested specimens, since 1 false-negative result was found, which was resolved on retesting with Sputasol pretreatment. Further studies need to determine the frequency of preprocessing of sputum samples before efficiently running the RP panel.
Specimens for inclusion in this study were previously tested at two different sites, using both their own systems and validated assays. Although the initial setups of the LDT assays were the same (11, 12), minor adjustments of the assays and the use of different PCR platforms may affect the performance of the LDTs and therefore were a limitation of this study.
Comparison of the results from the RP panel with the results from the routine multiplex real-time PCR showed an agreement of 97.4% in 464 pathogens tested. Figure 3 shows that PCR+/RP targets are mainly targets with a low viral or bacterial load (based on CT values). Analysis of real-time PCR results with CT values of <35 provided an agreement between the RP panel and laboratory-developed real-time PCR of 99.1%. Specimens containing pathogens with a high load (CT value of <30) were all detected correctly by the RP panel, independently of the type of specimen, type of pathogen, or the number of different pathogens in a specimen. This finding is in line with earlier evaluation of the GenMark XT-8 system using the same eSensor principle of detection (10). With 29 concordant targets with a CT value between 35 to 40 and 4 targets with a CT value of >40, the RP panel showed good detection rates with regard to lower viral or bacterial loads as well (Fig. 3).
Although the performance of the RP panel appeared to be excellent using the tested specimens in this study, for PIV4 (n = 2) and C. pneumoniae (n = 0) the number of clinical specimens that could be analyzed was too low for a proper assessment of the assay, which was a limitation of this study.
In 14 different specimens, the RP panel identified 15 pathogens that had not been detected by routine testing (PCR/RP+). In addition, one influenza A virus detected by LDT could be detected as influenza A H1N1 2009 virus by the RP panel. One of the selected negative samples was shown to contain an hAdV, while all other PCR/RP+ targets were detected as copathogens to other positive targets in the samples. All the PCR/RP+ targets were found in samples obtained from 1 institute. Discrepant analysis on the eluates of these samples was performed using the LDT of the other institute, where seven of the additionally detected pathogens could be confirmed by discrepant testing. These samples showed a relatively low viral load based on the mean CT value found (33.3), probably around the limit of detection of the initial LDT. It was unclear whether the 8 unresolved PCR/RP+ targets are false positive or the results of more efficient detection of multipathogen infections by the eSensor technology (10).
A small number of LDT-negative specimens (n = 20) was included in this study since the main objective of this study was to determine the performance of the RP panel in detecting respiratory pathogens. Although this is a limitation of the current study, we believe that this issue will be addressed extensively in upcoming prospective clinical studies.
Owing to the lack of clinical specimens containing MERS-CoV, dilutions of two different culture isolates were tested in this study, of which dilutions with CT values of <30 as shown by the laboratory-developed real-time PCR assay could be detected consistently. It should be noted that the real-time PCR assay has been developed for research use and has not yet been validated for clinical use.
Assessment of the RP panel using EQA samples from QCMD and Qnostics showed results that are in line with the results obtained from clinical specimens. A total of 4 targets included in the EQA samples could not be detected using the RP panel, showing CT values of >35 (n = 3) and 31.4 (n = 1) when tested by real-time PCR.
The RP panel on the ePlex system enables rapid testing and can be used as a diagnostic system in either a laboratory or a decentralized setting that is closer to the patient. The assay turned out to be rapid and straightforward to perform. Compared to routine testing, hands-on time of the RP panel was very low (<2 min), whereas the hands-on time of the routine testing was about 30 to 45 min, depending on the nature and number of samples tested. The overall run time of the platforms was also in favor of the ePlex system, as it takes approximately 90 min for nucleic acid extraction, amplification, hybridization, and detection, whereas routine testing takes up to 2 h and 45 min using different systems and multiple real-time PCR assays in multiplex. An important advantage of the ePlex system is the possibility of random access testing, compared to batch-wise testing in the current diagnostic real-time PCR approach. With a relatively short turnaround time and the potential to randomly load and run up to 24 specimens, the ePlex system is very suitable for testing STAT samples, which require immediate testing. In contrast to LDTs, where CT values represent a quantitative indicator, the ePlex system generates qualitative results only. The CT value is dependent on many different factors such as sample type and course of infection and can therefore differ greatly, even within a single patient. Hence, a qualitative result, e.g., identification of the pathogen, is the major factor for patient management.
The costs of reagents per sample are relatively high for ePlex compared to LDT. However, when taking into account the hands-on time of technicians and the clinical benefit of more rapid results, the assay will most likely be more cost-effective. Studies evaluating a rapid diagnostic assay for respiratory pathogens, such as the FilmArray respiratory panel (BioFire Diagnostics, Salt Lake City, UT), have already shown the impact of rapid diagnostics for respiratory pathogens, since it decreased the duration of antibiotic use, the length of hospitalization, and the time of isolation, delivering financial savings (13, 14). Although the RP panel on the ePlex system has the same potential, clinical studies remain to be conducted to fulfill this potential.
In conclusion, this study shows excellent performance of the GenMark ePlex RP panel in comparison to laboratory-developed real-time PCR assays for the detection of respiratory pathogens from multiple types of clinical specimens and EQA samples. The system provides a large amount of useful diagnostic data within a short time frame, with minimal hands-on time, helping to reduce laboratory costs for labor and deliver a faster result to the clinician in order to aid in appropriate antimicrobial therapy. Therefore, this syndrome-based diagnostic assay could be used as rapid diagnostic testing in many different settings.

MATERIALS AND METHODS

Specimens.

Clinical specimens selected for this study have previously been submitted and tested prospectively for diagnosis of respiratory infections at either the Specialist Virology Center at the Royal Infirmary of Edinburgh (RIE) or the medical microbiology laboratory at the Leiden University Medical Center (LUMC). Specimens were selected using the laboratory information management system of the corresponding institute, without prior selection based on CT value. Ethical approval for this study was granted by the medical ethical committee provided that anonymized samples were used. A total of 343 clinical specimens (286 nasopharyngeal [NPS] swabs, 21 sputum samples, 16 bronchoalveolar lavage [BAL] fluid samples, 10 throat swabs, and 10 nasopharyngeal aspirates) were used for this study, 323 positive and 20 negative. In the absence of clinical samples, Middle East respiratory syndrome coronavirus (MERS-CoV) isolate Jordan/N3 and recombinant MERS-CoV isolate EMC/2012 were tested. Finally, the QCMD 2016 Respiratory II Pilot study panel (12 samples) and a custom external quality assessment (EQA) evaluation panel of 17 samples provided by Qnostics Ltd. (Glasgow, United Kingdom) were tested.

Diagnostic testing by lab-developed tests.

In short, the routine testing method consisted of total nucleic acid extraction by the NucliSENS easyMAG system (∼45 min; bioMérieux, Basingstoke, United Kingdom) or the MagNA Pure LC system (∼45 min to 1 h 30 min depending on the number of samples; Roche Diagnostics, Almere, The Netherlands), at the RIE and the LUMC, respectively. An input volume of 200 μl per specimen and elution volume of 100 μl were used for all specimen types. Amplification and detection were performed by real-time PCR using the ABI 7500 fast thermocycler (1 h; Applied Biosystems, Paisley, United Kingdom) or the Bio-Rad CFX96 thermocycler (∼1 h 40 min; Bio-Rad, Veenendaal, The Netherlands), at the RIE and the LUMC, respectively. Real-time PCR assays were tested with updated versions (where needed) of primers and probes as described previously (11, 12).

RP panel.

Original clinical specimens were retrieved from storage at −70°C and thawed at room temperature. After vortexing, 200 μl of the specimen was pipetted into the sample delivery device with a buffer provided by the manufacturer. For 16 out of 21 sputum samples, preprocessing was done using Sputasol (Oxoid, Basingstoke, United Kingdom) according to the manufacturer's procedures (with the exception of washing the sputum) and incubation at 37°C for 15 min on a shaker at 500 rpm. After gentle mixing of the specimen and buffer in the sample delivery device, the mixture was dispensed into the cartridge using the sample delivery port, which was subsequently closed by sealing with a cap. After scanning of the barcode of the ePlex RP panel cartridge and the barcode of the corresponding sample, the cartridge was inserted into an available bay of the ePlex system. The test then started automatically and ran for approximately 90 min.
A single cartridge of the RP panel is able to detect 25 respiratory pathogens, including differentiation of subtypes of influenza A virus, parainfluenza virus, and respiratory syncytial virus (RSV) (Table 1). Internal controls for extraction, bead delivery, and movement within the cartridge are present, as well as those for amplification, digestion, and hybridization of DNA and RNA targets. For every specimen tested, a sample detection report was created, comprising the results for all targets and internal controls. Results of the targets are reported as positive or not detected. If an internal control fails, this will be noted on the detection report and samples should be retested with a new cartridge.

Discrepant testing.

In the case of discrepant results, the discordant sample was retested either with a new ePlex cartridge if the real-time PCR was positive and the RP panel was negative (PCR+/RP) or with the laboratory-developed real-time PCR assay in the case of PCR/RP+ results. For unresolved discrepancies, additional testing with a third PCR assay (different primers and probe) was performed for final resolution.

ACKNOWLEDGMENTS

We thank Yvette van Aarle, Mario Bussel, Wilfred Rijnsburger, and Tom Vreeswijk of the LUMC and Laura Mackenzie of the Edinburgh RIE for performing the assays including (discrepant) diagnostics. Also, we thank colleagues from the Global Emerging Infections Surveillance and Response System (GEIS) of the U.S. Naval Medical Research Unit 3 (NAMRU-3; Cairo, Egypt) for providing us the Jordan/N3 isolate and L. Enjuanes from the Centro Nacional de Biotecnologia (CNB-CSIC; Madrid, Spain) for providing the recombinant Middle East respiratory syndrome coronavirus isolate EMC/2012.
None of the authors have conflicts of interest to declare.
GenMark Diagnostics, Inc., provided kits and reagents to perform this study and was responsible for study design. GenMark Diagnostics, Inc., did not have any influence on the content of the submitted manuscript.

REFERENCES

1.
Torres A, Lee N, Cilloniz C, Vila J, and Van der Eerden M. 2016. Laboratory diagnosis of pneumonia in the molecular age. Eur Respir J48:1764–1778.
2.
Gaunt ER, Hardie A, Claas EC, Simmonds P, and Templeton KE. 2010. Epidemiology and clinical presentations of the four human coronaviruses 229E, HKU1, NL63, and OC43 detected over 3 years using a novel multiplex real-time PCR method. J Clin Microbiol48:2940–2947.
3.
Scheltinga SA, Templeton KE, Beersma MF, and Claas EC. 2005. Diagnosis of human metapneumovirus and rhinovirus in patients with respiratory tract infections by an internally controlled multiplex real-time RNA PCR. J Clin Virol33:306–311.
4.
Templeton KE, Scheltinga SA, Beersma MF, Kroes AC, and Claas EC. 2004. Rapid and sensitive method using multiplex real-time PCR for diagnosis of infections by influenza a and influenza B viruses, respiratory syncytial virus, and parainfluenza viruses 1, 2, 3, and 4. J Clin Microbiol42:1564–1569.
5.
Templeton KE, Scheltinga SA, Graffelman AW, Van Schie JM, Crielaard JW, Sillekens P, Van Den Broek PJ, Goossens H, Beersma MF, and Claas EC. 2003. Comparison and evaluation of real-time PCR, real-time nucleic acid sequence-based amplification, conventional PCR, and serology for diagnosis of Mycoplasma pneumoniae. J Clin Microbiol41:4366–4371.
6.
Templeton KE, Scheltinga SA, Sillekens P, Crielaard JW, van Dam AP, Goossens H, and Claas EC. 2003. Development and clinical evaluation of an internally controlled, single-tube multiplex real-time PCR assay for detection of Legionella pneumophila and other Legionella species. J Clin Microbiol41:4016–4021.
7.
Templeton KE, Scheltinga SA, van den Eeden WC, Graffelman AW, van den Broek PJ, and Claas EC. 2005. Improved diagnosis of the etiology of community-acquired pneumonia with real-time polymerase chain reaction. Clin Infect Dis41:345–351.
8.
Gadsby NJ, Hardie A, Claas EC, and Templeton KE. 2010. Comparison of the Luminex respiratory virus panel fast assay with in-house real-time PCR for respiratory viral infection diagnosis. J Clin Microbiol48:2213–2216.
9.
Bruijnesteijn van Coppenraet L, Swanink CMA, van Zwet AA, Nijhuis RHT, Schirm J, Wallinga JA, and Ruijs G. 2010. Comparison of two commercial molecular assays for simultaneous detection of respiratory viruses in clinical samples using two automatic electrophoresis detection systems. J Virol Methods169:188–192.
10.
Pierce VM and Hodinka RL. 2012. Comparison of the GenMark Diagnostics eSensor respiratory viral panel to real-time PCR for detection of respiratory viruses in children. J Clin Microbiol50:3458–3465.
11.
Loens K, van Loon AM, Coenjaerts F, van Aarle Y, Goossens H, Wallace P, Claas EJ, and Ieven M. 2012. Performance of different mono- and multiplex nucleic acid amplification tests on a multipathogen external quality assessment panel. J Clin Microbiol50:977–987.
12.
Templeton KE, Scheltinga SA, van der Zee A, Diederen BM, van Kruijssen A, Goossens H, Kuijper E, and Claas EC. 2003. Evaluation of real-time PCR for detection of and discrimination between Bordetella pertussis, Bordetella parapertussis, and Bordetella holmesii for clinical diagnosis. J Clin Microbiol41:4121–4126.
13.
Poelman R, van der Meer J, van Leer-Buter C, Riezebos-Brilman A, and Niesters HGM. 2015. Point-of-impact testing in the emergency department: rapid diagnostics for respiratory viral infections. J Clin Virol70(Suppl 1):S48.
14.
Rogers BB, Shankar P, Jerris RC, Kotzbauer D, Anderson EJ, Watson JR, O'Brien LA, Uwindatwa F, McNamara K, and Bost JE. 2015. Impact of a rapid respiratory panel test on patient outcomes. Arch Pathol Lab Med139:636–641.

Information & Contributors

Information

Published In

cover image Journal of Clinical Microbiology
Journal of Clinical Microbiology
Volume 55Number 6June 2017
Pages: 1938 - 1945
Editor: Alexander J. McAdam, Boston Children's Hospital
PubMed: 28404682

History

Received: 18 February 2017
Returned for modification: 8 March 2017
Accepted: 4 April 2017
Published online: 23 May 2017

Keywords

  1. influenza
  2. point-of-care testing
  3. rapid diagnostics
  4. respiratory pathogens
  5. sample-to-answer

Contributors

Authors

R. H. T. Nijhuis
Department of Medical Microbiology, Leiden University Medical Center, Leiden, The Netherlands
D. Guerendiain
Specialist Virology Center, Royal Infirmary of Edinburgh, Edinburgh, United Kingdom
E. C. J. Claas
Department of Medical Microbiology, Leiden University Medical Center, Leiden, The Netherlands
K. E. Templeton
Specialist Virology Center, Royal Infirmary of Edinburgh, Edinburgh, United Kingdom

Editor

Alexander J. McAdam
Editor
Boston Children's Hospital

Notes

Address correspondence to R. H. T. Nijhuis, [email protected].

Metrics & Citations

Metrics

Note:

  • For recently published articles, the TOTAL download count will appear as zero until a new month starts.
  • There is a 3- to 4-day delay in article usage, so article usage will not appear immediately after publication.
  • Citation counts come from the Crossref Cited by service.

Citations

If you have the appropriate software installed, you can download article citation data to the citation manager of your choice. For an editable text file, please select Medlars format which will download as a .txt file. Simply select your manager software from the list below and click Download.

View Options

Figures

Tables

Media

Share

Share

Share the article link

Share with email

Email a colleague

Share on social media

American Society for Microbiology ("ASM") is committed to maintaining your confidence and trust with respect to the information we collect from you on websites owned and operated by ASM ("ASM Web Sites") and other sources. This Privacy Policy sets forth the information we collect about you, how we use this information and the choices you have about how we use such information.
FIND OUT MORE about the privacy policy